MSDD1000 series Monochromator–spectrograph is a double dispersion monochromator-spectrograph with the effective focal length of 1 m. Double scheme combines a high spectral resolution which is peculiar to long-focus instruments, extremely low stray light being the advantage of double schemes and small sizes of single monochromators–spectrographs.

MSDD1000 series Monochromator–spectrograph is a completely automated spectral instrument with the effective focal length of 1000 mm (2х500 mm) and F\number of 1/5.9, which is applicable as a monochromator or a spectrograph with a large flat field.

 

The MSDD1000 series is designed on the modified Cherny-Turner scheme with the two-level arrangement of optical components to allow dispersion doubling, aberrations minimization and exclusion of spectra re-entrance from the diffraction gratings and mirrors.

This is a double dispersion monochromator with no intermediate slit, in which upper and lower levels are the separate monochromators. These two levels are connected by means of two plane turning mirrors.
The design of the optical scheme and construction of the MSDD1000 series allows the monochromator of any level to be applied as the first one. Thus, if radiation is delivered to the entry of monochromator of one of the levels, the spectrum is formed in the flat field of the monochromator of another level, providing easy cascade of monochromators for construction of quadruple monochromator from two double ones.

High-quality optics and a wide choice of diffraction gratings provide operation of the MSDD1000 series monochromator-spectrograph in a wide spectral range from 185 nm up to 60 µm (depending on the selected grating) with the best spectral resolution.

In MSDD1002i model and MSDD1004i model the astigmatism is corrected owing to special optics what makes these instruments applicable in multi-channel spectroscopy. ONLY one input port is used in these instruments.

 

MSDD1000 series Monochromator–spectrograph has two input (only the models with no astigmatism compensation - MSDD1002 model and MSDD1004 model) ports and two output ports. Selection of the output ports is realized with flip mirrors. Either exit spectral slits or multi-channel detectors can be established in each of the output ports.

MSDD1000 series Monochromator–spectrograph has two types of entrance and exit spectral slits: manual, operated with a micrometer screw, and automated (combined) when the slit width is operated automatically or manually (with micrometer).

 

A wide choice of changeable gratings gives the best combination of high energetic efficiency and maximal spectral resolution for different spectrum areas.

We also propose a great choice of additional accessories for MSDD1000 series monochromator–spectrograph: filter wheel (6 and 8 filters), fast shutters with the opportunity of radiation direct delivery or via optical fiber, optical fibers with adapters, input and output condensers, integral and multi-element detectors and etc.

Extremely high spatial resolution (better than 0,004 nm for 3600l/mm grating) in combination with high throughput (F/number 1/5.9) make double dispersion monochromator–spectrographs of MSDD1000 series an ideal instrument for spectroscopy applications requiring high resolution in the limited area, when other instruments can not be physically applied.

 

I M A G I N G
 

SOL instruments: спектрометр, рамановский микроскоп, эмиссионный спектрометр. msdd1004i photo

 

MS benefits

 

MSDD1004 optic sheme

Automated turret for two pairs of gratings (MSDD1004 model and MSDD1004i model) provides a fast switching of gratings with high repeatability of the set wavelength.

 

MSDD1002 optic sheme

In MSDD1002 model and MSDD1002i model a fixed holder is used for easy manual switching of gratings.

 

SOL instruments: спектрометр, рамановский микроскоп, эмиссионный спектрометр. book-2

Theory & Instructions
1. Spectral Instrument. Basic Concepts and Characteristics.
2. Spectral Devices. A Spectral Device Selection. 
3. Spectra Detection. Detector Selection.

 

  

Models MSDD100 Series MSDD1002 MSDD1002i MSDD1004 MSDD1004i
Diffraction unit Fixed holder for manually changeable gratings Fixed holder for manually changeable gratings Turret for two pairs of gratings Turret for two pairs of gratings
Optics  Standard Astigmatism–compensated (Imaging) Standard Astigmatism–compensated (Imaging)

 

As well as in all Cherny–Turner scheme–based spectral instruments, in MSDD1000 series monochromator–spectrographs an 00' axis of the diffraction gratings rotation goes through the center of grating operating surface and coincides with the direction of lines. This fact provides the stability of the geometry of the beams for formation of a plane and fixed flat field. In the instrument with an automated turret, gratings rotation axes M-M1 and N-N1 are arranged perpendicular to the spectrum scanning axis, providing a high repeatability of wavelength setting at replacing of gratings.

產品規格

Optical system

Optical scheme: modified Cherny–Turner two-fold configuration
Wavelengths range: 180 nm - 60 µm
F\number: 5.9
Focal length: 1000 mm (2 х 500 mm)
Scanning range limited by the grating rotation angle: 0 - 1280 nm (grating 1200 l/mm)
Basic mirrors: Spherical
Stray light: 10-8 (20 nm from laser line 632,8 nm)
Flat field: 28 х 10 mm

Imaging

Model MSDD1002 MSDD1004 MSDD1002i MSDD1004i
Horizontal magnification: 1.08 1.08 1.08 1.08
Vertical magnification: - - 1.11 1.11
Vertical spatial resolution
Minimal distance between the edges of two neighboring tracks (in multi-channel spectroscopy):
- - < 15 µm < 15 µm

 

Spectral Image Enlarged Spectral Image

Spectral image in the focal plane of MSDD1004i.
Grating 1200 l/mm, light source - mercury-helium lamp: (576.961 nm and 579.067 nm), multi-channel optical fiber bundle, 
fiber core diameter 200 µm, CCD pixel size 12x12 µm)

Drive

Motor: micro stepper
Drive: sine
Step size: 0.18 arc seconds
Precision: ± 1 step
Max. speed: 10000 steps/s

Optic (for grating 1200 l/mm, wavelength 546 mm)

Reciprocal dispersion: 0.82 nm/mm
Repeatability: ± 0.005 nm
Spectral resolution:  
- PMT: 0.013 nm
- digital cameras (12 µm pixel): 0.017 nm
Wavelength accuracy: ± 0.028 nm
Average scanning step: 0.0014 nm

Gratings

Size: 80 х 70 х 10 mm
Rotation: About the center of ruled grating surface
Mounting: 1. Automated turret for 2 pairs of gratings
2. Holder for 1 pair of manually changeable grating
Automated turret for 2 pairs of gratings (MSDD1004 model and MSDD1004i model)
Grating repeatability:
 
- wavelength:

± 0.006 nm

- vertical image:

± 20 µm

Holder for 1 manually changeable grating (models MSDD1002 and MSDD1002i)
Grating repeatability:

 

 

- wavelength: ± 0.05 nm
- vertical image: ± 75 µm

Ports

Numbers: 2 input, 2 output
Output port repeatability (automated mirror): ± 0.01 nm
Switching time (input port): 4 s
Switching time (output port): 4 s

Spectral slits automated (combined)

Control: Automated (drive) or manual (micrometer)
Width: Regulated from 0 up to 2.0 mm
Parallelism: ± 1 µm
Accuracy (slit 1 mm): ± 10 µm
Repeatability: ± 1 µm
Reading accuracy (micrometer): 2 µm
Step size: 0.5 µm
Height: Diaphragm-regulated from 0 up to 10 mm

 
Spectral slits (manually operated)

Control: Manual (micrometer)
Width: Regulated from 0 up to 2.0 mm
Parallelism: ± 1 µm
Accuracy (slit 1 mm): ± 10 µm
Repeatability: ± 1.5 µm
Reading accuracy (micrometer): 2 µm
Height: Diaphragm-regulated from 0 up to 10 mm

Integrated shutter

Number: 2 (in each input port)
Shut time: ~ 100 ms
Max. frequency: 1 Hz
Control: On-board CPU or TTL-signals from the external device

Control

Central: On-board CPU

External:

PC
External interface: Ethernet

Dimensions

Overall sizes (LхWхH): 556 х 324 х 340 mm

Weight:

35 kg (may differ, depends on configuration)

 